Lasting accuracy for diaphragm seal systems for vacuum process

A diaphragm seal system in vacuum pressure process must master a major challenge ? it must deliver a trusted measuring result, even when the pressure drops and the temperature rises as well. The example of a chemical company producing phthalic anhydride (PA) demonstrates it is by no means self-evident that requirement could be met on a frequent basis.
PA is primarily used as a plasticiser for plastics. However, additionally it is found in the production of synthetic resins and dyes. The plant operator obtains this compound via the catalytic oxidation of o-xylene in the gas phase, followed by vacuum pressure distillation process.
Critical phase for pressure measurement
In this process, the critical phase for pressure measurement occurs in the distillation column. There, Expert can rise to more than 230 �C while the pressure drops to 100 mbar abs. With such a contrary action, the conventional installed measurement technology reaches its performance limit relatively quickly, that your aforementioned manufacturer was forced to find out ? all instrumentation showed a measurement deviation beyond your defined tolerance after only a fortnight.
Diaphragm seal system adapted in all elements to the vacuum process
The diaphragm seal system is tailored exactly to the conditions of the vacuum process and tested under application conditions.
Consequently, the company approached WIKA with the request for a measuring solution that would also fulfil its task with the mandatory accuracy in the long run under the conditions described. WIKA then drew on its in-depth expertise to design a diaphragm seal system for the vacuum process, comprising an activity transmitter (model IPT-20), a flange-type diaphragm seal with flush diaphragm (model 990.27) and a capillary.
All components of this measuring system were selected specifically for the application. Silicone oil KN32 was therefore selected because the fill fluid for pressure transmission. The cleaning and conditioning of the diaphragm seal, the capillary, the bond pieces as well as the automated filling process were also matched to each other exactly. With Today , even one drop of liquid too much or inadequate can destabilise the quantity balance of the diaphragm seal system and therefore impair its function. Because of this, correct measurement would no longer be guaranteed.
Test under application conditions
After cleaning, all parts of the diaphragm seal system were first fully welded right into a single unit as usual, accompanied by the filling and adjustment of the machine. Before delivery, however, there is another crucial step ? in its laboratory, WIKA simulated vacuum pressure process for the diaphragm seal system in accordance with the client?s conditions and subjected it to a thermal ageing process. This test confirmed the procedure suitability of the measuring system.
Note
Further information on the WIKA diaphragm seal systems can be found on the WIKA website. If you have any questions, your contact will gladly assist you to.